ServicenavigationHauptnavigationTrailKarteikarten


Forschungsstelle
EU FRP
Projektnummer
96.0438
Projekttitel
Dosimetry for radiation sterilization of medical devices
Projekttitel Englisch
Dosimetry for radiation sterilization of medical devices

Texte zu diesem Projekt

 DeutschFranzösischItalienischEnglisch
Schlüsselwörter
-
-
-
Anzeigen
Alternative Projektnummern
-
-
-
Anzeigen
Forschungsprogramme
-
-
-
Anzeigen
Kurzbeschreibung
-
-
-
Anzeigen
Partner und Internationale Organisationen
-
-
-
Anzeigen
Abstract
-
-
-
Anzeigen
Datenbankreferenzen
-
-
-
Anzeigen

Erfasste Texte


KategorieText
Schlüsselwörter
(Englisch)
Radiation sterilization; dosimetry; calibration; medical device
Alternative Projektnummern
(Englisch)
EU project number: SMT4-CT96-2077
Forschungsprogramme
(Englisch)
EU-programme: 4. Frame Research Programme - 2.2 Measurements and testing
Kurzbeschreibung
(Englisch)
See abstract
Partner und Internationale Organisationen
(Englisch)
NPL England, Inst. Riso DK
Abstract
(Englisch)
The purpose of the project is to give a common standard for the correct application of high- Dose Dosimetry for the industrial sterilisation of medical devices.
During a first intercomparison between 27 cobalt 60 plants and 11 electron beam facilities with 2 dosimetry systems (alanine and dichromate) the measured doses were compared. It showed that a remarkable deviation between the different plants occurred. Following this first intercomparison a draft of a calibration guideline was developed, who should help to improve accuracy of calibration.
In a second intercomparison under equal conditions as the first one was then performed, following the guideline draft. The results of the second intercomparison showed, that the deviations between the different participants could be significantly reduced and more than 90% of the participants showed an agreement of +/- 5% for gamma and +/- 12% for electron beam treatment.
Therefore the developed guideline for calibration of dosimeters leads to more accurate and reliable dosimetry readings and helps to assure a precise and safe application of irradiation for the sterilisation of medical devices.