METAS adapted and characterized their dynamic generation systems to produce NO2 reference gas mixtures at low amount fractions (10 – 500 nmol/mol). By modifying the protocol for the generation and extending the calibration period to 6 days instead of 3, the NO2 reference gas mixtures were able to be produced with a relative expanded uncertainty (k = 2) of 0.9 % (1-step dilution) and 1.0 % (2-step dilution). The NO2 reference gas mixtures generated using ReGaS1 had a relative expanded uncertainty (k = 2) between 2.5 % and 2.7 % for 1- and 2-step dilutions.
In addition, the dynamic generation systems were used to calibrate and generate HNO3 reference gas mixtures that were used in activities A2.2.2, A3.3.1, and A3.3.3. For the first time, stable and SI-traceable HNO3 reference gas mixtures were generated, although with associated uncertainties (> 30%) above the target uncertainty (< 5%). The main contribution to the uncertainty (99.9 %) was the purity of the HNO3 permeation tubes.
5 different permeators (3 NO2 permeators and 3 HNO3 permeators) were tested for temporal stability and the effects of temperature, pressure, and preconditioning.
- Temporal stability: For NO2 the permeation rate remained stable in the first 20 months, but increased in month 25, by > 3%, and by up to 7% by month 30. HNO3 permeators were evaluated over 24 months, with no clear trend.
- Temperature: An increase of 1 oC resulted in a 4% and 8% increase in permeation rates for NO2 and HNO3, respectively.
- Pressure: Permeation rates decreased linearly with increasing pressure, however, pressure variations during the calibration of a permeator does not affect the permeation rate. For pressures varying between 800 mbar and 2600 mbar, the permeation rate remained within the uncertainty.
- Pre-conditioning: The time needed for the membrane to reach stability was 3000 minutes for NO2 and 1200 minutes for HNO3, which has a thicker membrane. This period was reduced by keeping the permeator at calibration conditions (temperature and pressure) for 3 – 5 days before placing the permeator in MSB.
Further optimization of the FTIR-spectrometer analytical system is needed in order to detect and measure HNO3 impurities in NO2 permeators. HNO3 coming out of the permeator were lower than the detection limit of the FTIR.
A field-based side-by-side comparison organized by Deutsche Wetter Dienst in FZ-Jülich included calibrated NO2 and HNO3 with the portable generator ReGaS1 to calibrate the measurement instruments. Results from the comparison were included in deliverable D7: "Report on the field-based side-by-side comparison of selective NO2 instrumentation", which will be submitted as a manuscript by FZ-Jülich (April 2021), and co-authored by METAS.